qtOCT: quantitative transmission optical coherence tomography (2024)

Wojciech Krauze\authormark   Martyna Mazur   and Arkadiusz KuśWarsaw University of Technology, Institute of Micromechanics and Photonics, Boboli 8 street, 02-525, Warsaw Poland\authormark*wojciech.krauze@pw.edu.pl

journal: opticajournalarticletype: Research Article

{abstract*}

Transmission optical coherence tomography (OCT) enables analysis of biological specimens in vitro through detection of forward scattered light. Up to now, transmission OCT was considered as a technique that cannot directly retrieve quantitative phase and is thus a qualitative method. In this paper, we present qtOCT, a novel quantitative transmission optical coherence tomography method. Unlike existing approaches, qtOCT allows for a direct, easy, fast and rigorous retrieval of 2D integrated phase information from transmission full-field swept-source OCT measurements. Our method is based on coherence gating and allows user-defined temporal measurement range selection, making it potentially suitable for analyzing multiple-scattering samples. We demonstrate high consistency between qtOCT and digital holographic microscopy phase images. This approach enhances transmission OCT capabilities, positioning it as a viable alternative to quantitative phase imaging techniques.

1 Introduction

Optical coherence tomography (OCT) is an imaging technique, most commonly realized in reflection configuration where the sample is illuminated in the epi mode [1]. It allows retrieving qualitative information about refractive index gradients with relatively high axial resolution in vivo, however with refraction-induced distortions. In recent years, transmission OCT (tOCT) systems have gained attention of the scientific community [2]. One reason behind this interest is that in most biological samples forward scattered light is much stronger than the backward scattered light due to their high anisotropy parameter [3]. Additionally, lack of epi mode measurement is not an issue when biological in vitro measurements are carried out. Unfortunately, despite the fact that the theoretical optical transfer function in tOCT systems is in the low frequency region (covering the DC term) [4], up to now there is no direct method that can retrieve the phase from such OCT measurements. Indirect methods include localization of the ballistic peak position in point scanning OCT [5, 6], which requires very precise control of the optical path difference (OPD) between sample and reference arms of an OCT system and can be applied only where the ballistic peak is clearly visible and thus are not universal and practical. Another indirect approach is based on holoscopy-like systems [7, 8] which do not take advantage of the OCT imaging properties, like high sensitivity due to cross-correlation of the sample and reference beams, and instead process captured data in a classical interferometry approach. Anna et al. [9] presented a method of retrieving phase images from tOCT images, however it was implemented erroneously, which is confirmed by wrong phase values in the reconstructed images.

In this paper we present the first to our knowledge direct method to retrieve complex amplitudes from measurements captured with full-field swept-source OCT in transmission (t-FF-SS-OCT). Our method, called qtOCT (quantitative transmission OCT), is a novel signal processing method strictly based on OCT theory and as such it does not require any modifications to standard t-FF-SS-OCT systems. We identify 3 main advantages of qtOCT. Firstly, this is a first method where 2D transverse quantitative phase information can be directly retrieved from tOCT data. In that sense, the resulting information can also be referred to as quantitative, integrated en face image. Secondly, since our solution is based on coherence gating phenomenon, it allows straightforward selection of user-defined temporal measurement range, which can be used to discard multiple scattered photons and reduce the problem to a weakly scattering one. What is also important, qtOCT is relatively easy to implement: for those who already have a digital holographic microscopy (DHM) system [10], it is enough to change the light source to a swept-source laser, and for those who have the t-FF-SS-OCT setup it is enough to change the signal processing algorithm.

The paper is structured as follows. In Section 2 we present the details of our method and explain important differences in interpretation of tOCT measurements in comparison to traditional reflection configuration, which have profound consequences. In Section 3 we present and discuss the measurement results that prove the quantitative nature of qtOCT and its temporal gating property, and finally in Section 4 we give the conclusions.

2 Method

2.1 Optical setup

The measurement system used in this work is an off-axis DHM based on Mach-Zehnder interferometer (MZI) setup (Fig.1). The light source is a swept-source (BS-840-1-HP, Superlum) with the full wavelength range of the source ΔλΔ𝜆\Delta\lambdaroman_Δ italic_λ of 75nmtimes75nanometer75\text{\,}\mathrm{nm}start_ARG 75 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG starting from 803nmtimes803nanometer803\text{\,}\mathrm{nm}start_ARG 803 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG to 878nmtimes878nanometer878\text{\,}\mathrm{nm}start_ARG 878 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG at 20mWtimes20milliwatt20\text{\,}\mathrm{mW}start_ARG 20 end_ARG start_ARG times end_ARG start_ARG roman_mW end_ARG power. The linewidth of the source is 0.06nmtimes0.06nanometer0.06\text{\,}\mathrm{nm}start_ARG 0.06 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG, which results in a coherence length of 10.7mmtimes10.7millimeter10.7\text{\,}\mathrm{mm}start_ARG 10.7 end_ARG start_ARG times end_ARG start_ARG roman_mm end_ARG at λ=803nm𝜆times803nanometer\lambda=$803\text{\,}\mathrm{nm}$italic_λ = start_ARG 803 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG. We use a mirror system in the object beam and a cat-eye retro-reflector (CE in Fig.1) in the reference beam, to have full control over the OPD in the system. The sample is illuminated by a beam diameter of 220µmtimes220micrometer220\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 220 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG (at 1/e21superscript𝑒21/e^{2}1 / italic_e start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT) and imaged by a 40x40𝑥40x40 italic_x NA 1.31.31.31.3 microscope objective (Zeiss) and a tube lens (effective focal length EFL=300µmtimes300micrometer300\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 300 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG), with magnification of M=71.775M71.775\text{M}=-71.775M = - 71.775 (measured separately). The beam expander (BE) is used to match the diameter of the reference beam to the object beam at the detector plane. The detector used in this system is a monochromatic CMOS camera (acA2040-180-km, Basler) with 5.5µmtimes5.5micrometer5.5\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 5.5 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG pixels arranged in a 2048x20482048𝑥20482048x20482048 italic_x 2048 array, offering a maximum framerate of 180180180180 fps, which puts the limit on the measurement speed as the source offers a 10 000nm/stimes10000nms10\,000\text{\,}\mathrm{n}\mathrm{m}\mathrm{/}\mathrm{s}start_ARG 10 000 end_ARG start_ARG times end_ARG start_ARG roman_nm / roman_s end_ARG scan speed, which translates to 133.3133.3133.3133.3 full sweeps per seconds.

qtOCT: quantitative transmission optical coherence tomography (1)

The swept-source provides nearly flat spectral power distribution, thus axial temporal resolution can be defined as δt=λc2/(Δλc)=31fssubscript𝛿𝑡superscriptsubscript𝜆𝑐2Δ𝜆𝑐times31femtosecond\delta_{t}=\lambda_{c}^{2}/(\Delta\lambda c)=$31\text{\,}\mathrm{fs}$italic_δ start_POSTSUBSCRIPT italic_t end_POSTSUBSCRIPT = italic_λ start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT / ( roman_Δ italic_λ italic_c ) = start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG. Here, λcsubscript𝜆𝑐\lambda_{c}italic_λ start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT and c𝑐citalic_c indicate center wavelength and speed of light, respectively. The lateral resolution is 0.32µmtimes0.32micrometer0.32\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 0.32 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG.

2.2 Signal processing

qtOCT: quantitative transmission optical coherence tomography (2)

Two t-FF-SS-OCT datasets are captured for Nksubscript𝑁𝑘N_{k}italic_N start_POSTSUBSCRIPT italic_k end_POSTSUBSCRIPT wavenumbers: one object and one reference measurement. After that, qtOCT processing is applied. Graphical visualization of the processing pipeline is shown in Fig. 2A. The tOCT signal captured for each wavenumber k𝑘kitalic_k is a real-valued signal in the form of spectral interference fringes and can be described by [11]:

I(k)=κrefI0(k)+κobI0(k)FTtk{Φ(ct)Φ(ct)}++κobκrefI0(k)FTtk{Φ(ct)}+κobκrefI0(k)FTtk{Φ(ct)}𝐼𝑘subscript𝜅refsubscript𝐼0𝑘subscript𝜅obsubscript𝐼0𝑘superscriptFTsuperscript𝑡𝑘Φ𝑐superscript𝑡superscriptΦ𝑐superscript𝑡subscript𝜅obsubscript𝜅refsubscript𝐼0𝑘superscriptFT𝑡𝑘Φ𝑐𝑡subscript𝜅obsubscript𝜅refsubscript𝐼0𝑘superscriptFT𝑡𝑘superscriptΦ𝑐𝑡\begin{split}I(k)&=\kappa_{\text{ref}}I_{0}(k)+\kappa_{\text{ob}}I_{0}(k)\text%{FT}^{t^{\prime}\rightarrow k}\{\Phi(ct^{\prime})\circledast\Phi^{*}(ct^{%\prime})\}+\\&+\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(k)\text{FT}^{t\rightarrow k%}\{\Phi(ct)\}+\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(k)\text{FT}^{t%\rightarrow k}\{\Phi(ct)\}^{*}\end{split}start_ROW start_CELL italic_I ( italic_k ) end_CELL start_CELL = italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_k ) + italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_k ) FT start_POSTSUPERSCRIPT italic_t start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT ) ⊛ roman_Φ start_POSTSUPERSCRIPT ∗ end_POSTSUPERSCRIPT ( italic_c italic_t start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT ) } + end_CELL end_ROW start_ROW start_CELL end_CELL start_CELL + square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_k ) FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t ) } + square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_k ) FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t ) } start_POSTSUPERSCRIPT ∗ end_POSTSUPERSCRIPT end_CELL end_ROW(1)

where I0(k)subscript𝐼0𝑘I_{0}(k)italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_k ) is a light source spectrum, κrefsubscript𝜅ref\kappa_{\text{ref}}italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT, κobsubscript𝜅ob\kappa_{\text{ob}}italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT represent spectral scaling factors for the reference and object arms, respectively, c𝑐citalic_c is a speed of light, tsuperscript𝑡t^{\prime}italic_t start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT characterize transmission time of the sample signal and t𝑡titalic_t is the time delay between reference and sample signals. The operators FT𝐹𝑇FTitalic_F italic_T, \circledast, represent Fourier transform, convolution and complex conjugate, respectively.

In order to retrieve information about investigated sample’s integrated phase Φ(ct)Φ𝑐𝑡\Phi(ct)roman_Φ ( italic_c italic_t ), one-dimensional (1D) Fourier transform along k𝑘kitalic_k is performed. The result is a standard OCT reconstruction:

IFTkt{I(k)}=κrefI0(ct)+κobI0(ct)ACF{Φ(ct)}++κobκrefI0(ct)Φ(ct)+κobκrefI0(ct)Φ(ct)superscriptIFT𝑘𝑡𝐼𝑘subscript𝜅refsubscript𝐼0𝑐𝑡subscript𝜅obsubscript𝐼0𝑐𝑡ACFΦ𝑐superscript𝑡subscript𝜅obsubscript𝜅refsubscript𝐼0𝑐𝑡Φ𝑐𝑡subscript𝜅obsubscript𝜅refsubscript𝐼0𝑐𝑡Φ𝑐𝑡\begin{split}\text{IFT}^{k\rightarrow t}\{I(k)\}&=\kappa_{\text{ref}}I_{0}(ct)%+\kappa_{\text{ob}}I_{0}(ct)\text{ACF}\{\Phi(ct^{\prime})\}+\\&+\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(ct)\circledast\Phi(ct)+%\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(ct)\circledast\Phi(-ct)\end{split}start_ROW start_CELL IFT start_POSTSUPERSCRIPT italic_k → italic_t end_POSTSUPERSCRIPT { italic_I ( italic_k ) } end_CELL start_CELL = italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) + italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) ACF { roman_Φ ( italic_c italic_t start_POSTSUPERSCRIPT ′ end_POSTSUPERSCRIPT ) } + end_CELL end_ROW start_ROW start_CELL end_CELL start_CELL + square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) ⊛ roman_Φ ( italic_c italic_t ) + square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) ⊛ roman_Φ ( - italic_c italic_t ) end_CELL end_ROW(2)

where IFT is an inverse Fourier transform, I0(ct)subscript𝐼0𝑐𝑡I_{0}(ct)italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) is a coherence function of the light source and ACF denotes autocorrelation.

The first 2 components in Eq.2 are DC and autocorrelation terms, and the information about Φ(ct)Φ𝑐𝑡\Phi(ct)roman_Φ ( italic_c italic_t ) is encoded in the last two complex conjugate cross-correlation components. To select one of them, a window function (CC window) is applied. After windowing, we receive Φ(ct)Φ𝑐𝑡\Phi(ct)roman_Φ ( italic_c italic_t ) convolved with I0(ct)subscript𝐼0𝑐𝑡I_{0}(ct)italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ):

CCobj=κobκrefI0(ct)Φ(ct)subscriptCCobjsubscript𝜅obsubscript𝜅refsubscript𝐼0𝑐𝑡Φ𝑐𝑡\text{CC}_{\text{obj}}=\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(ct)%\circledast\Phi(ct)CC start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT = square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) ⊛ roman_Φ ( italic_c italic_t )(3)

The convolution operation leads to broadening and distortion of Φ(ct)Φ𝑐𝑡\Phi(ct)roman_Φ ( italic_c italic_t ) signal (see blue box in Fig.2A) which decreases achievable resolution and accuracy. To decouple the OCT signal from the light source properties, 1D FT of CCobjsubscriptCCobj\text{CC}_{\text{obj}}CC start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT is calculated:

Uobj=FTtk{CCobj}=FTtk{κobκrefI0(ct)}FTtk{Φ(ct)}subscriptUobjsuperscriptFT𝑡𝑘𝐶subscript𝐶objsuperscriptFT𝑡𝑘subscript𝜅obsubscript𝜅refsubscript𝐼0𝑐𝑡superscriptFT𝑡𝑘Φ𝑐𝑡\text{U}_{\text{obj}}=\text{FT}^{t\rightarrow k}\{CC_{\text{obj}}\}=\text{FT}^%{t\rightarrow k}\{\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(ct)\}\cdot%\text{FT}^{t\rightarrow k}\{\Phi(ct)\}U start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT = FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { italic_C italic_C start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT } = FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) } ⋅ FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t ) }(4)

and the result UobjsubscriptUobj\text{U}_{\text{obj}}U start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT is divided by the reference measurement UrefsubscriptUref\text{U}_{\text{ref}}U start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT, which underwent the same signal processing as the object measurement:

Uref=FTtk{CCref}=FTtk{κobκrefI0(ct)}subscriptUrefsuperscriptFT𝑡𝑘𝐶subscript𝐶refsuperscriptFT𝑡𝑘subscript𝜅obsubscript𝜅refsubscript𝐼0𝑐𝑡\text{U}_{\text{ref}}=\text{FT}^{t\rightarrow k}\{CC_{\text{ref}}\}=\text{FT}^%{t\rightarrow k}\{\sqrt{\kappa_{\text{ob}}\kappa_{\text{ref}}}I_{0}(ct)\}U start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT = FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { italic_C italic_C start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT } = FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { square-root start_ARG italic_κ start_POSTSUBSCRIPT ob end_POSTSUBSCRIPT italic_κ start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG italic_I start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT ( italic_c italic_t ) }(5)
FTtk{Φ(ct)}=UobjUref=FTtk{CCobj}FTtk{CCref}superscriptFT𝑡𝑘Φ𝑐𝑡subscriptUobjsubscriptUrefsuperscriptFT𝑡𝑘𝐶subscript𝐶objsuperscriptFT𝑡𝑘𝐶subscript𝐶ref\text{FT}^{t\rightarrow k}\{\Phi(ct)\}=\frac{\text{U}_{\text{obj}}}{\text{U}_{%\text{ref}}}=\frac{\text{FT}^{t\rightarrow k}\{CC_{\text{obj}}\}}{\text{FT}^{t%\rightarrow k}\{CC_{\text{ref}}\}}FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t ) } = divide start_ARG U start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT end_ARG start_ARG U start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT end_ARG = divide start_ARG FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { italic_C italic_C start_POSTSUBSCRIPT obj end_POSTSUBSCRIPT } end_ARG start_ARG FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { italic_C italic_C start_POSTSUBSCRIPT ref end_POSTSUBSCRIPT } end_ARG(6)

Then, we calculate 1D IFT to return to the OCT reconstruction space. Due to performed source-correction, the broadening effect is suppressed and the resolution is significantly increased (see blue and orange boxes in Fig.2A). At this point, the CC window is corrected to keep only the signals for which delays relative to the reference signal are within selected period of time ΔtΔ𝑡\Delta troman_Δ italic_t, enabling temporal gating property. After that, 1D FT is performed, and finally integrated complex amplitudes for different k𝑘kitalic_k are obtained:

E(r)(k)=exp(i(kt))FTtk{Φ(ct)}superscript𝐸𝑟𝑘exp𝑖𝑘𝑡superscriptFT𝑡𝑘Φ𝑐𝑡E^{(r)}(k)=\text{exp}(i(kt))\cdot\text{FT}^{t\rightarrow k}\{\Phi(ct)\}italic_E start_POSTSUPERSCRIPT ( italic_r ) end_POSTSUPERSCRIPT ( italic_k ) = exp ( italic_i ( italic_k italic_t ) ) ⋅ FT start_POSTSUPERSCRIPT italic_t → italic_k end_POSTSUPERSCRIPT { roman_Φ ( italic_c italic_t ) }(7)

It is now possible to select E(r)superscript𝐸𝑟E^{(r)}italic_E start_POSTSUPERSCRIPT ( italic_r ) end_POSTSUPERSCRIPT for a specific k𝑘kitalic_k and retrieve integrated phase of a sample with standard approach by using two-argument arctangent function and a 2D phase unwrapping algorithm [12].

To give more insight into the operating principle of qtOCT and to show critical differences between transmission and reflection OCT modalities, two examples representing different investigated samples are shown in Fig.2. As a first example, a flat plate with two surfaces located at different depths z1subscript𝑧1z_{1}italic_z start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT and z2subscript𝑧2z_{2}italic_z start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT was chosen (see Fig.2B). Starting the consideration with the OCT signal formation in reflection mode, the incident light travelling trough the sample is backreflected from z1subscript𝑧1z_{1}italic_z start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT and z2subscript𝑧2z_{2}italic_z start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT surfaces. Due to the fact that z1<z2subscript𝑧1subscript𝑧2z_{1}<z_{2}italic_z start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT < italic_z start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT, the light from z2subscript𝑧2z_{2}italic_z start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT surface will reach the detector with a delay of ΔtΔ𝑡\Delta troman_Δ italic_t in relation to light coming from z1subscript𝑧1z_{1}italic_z start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT. The obtained ΔtΔ𝑡\Delta troman_Δ italic_t is here determined by the relative location of sample surfaces. This means that in the OCT reconstruction we will see two distinct layers directly corresponding to the depth structure of the flat plate. The situation is different in transmission OCT. The presence of two layers z1subscript𝑧1z_{1}italic_z start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT and z2subscript𝑧2z_{2}italic_z start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT does not generate time delays (Δt=0Δ𝑡0\Delta t=0roman_Δ italic_t = 0) in transmitted light. The captured signal represents the cumulative information from two surfaces rather than separate information from each of them, resulting in a single layer in the OCT reconstruction. Thus, in transmission OCT, and qtOCT specifically, in such case there is no sectioning property. On the other hand, this also means that qtOCT gives a unique opportunity to provide quantitative integrated phase, similarly to techniques like DHM.

The second example depicted in Fig.2C further explains the concept of the OCT signal generation in transmission mode. A hom*ogeneous microsphere immersed in a highly scattering medium is now an investigated sample. The object and medium have refractive indices n1subscript𝑛1n_{1}italic_n start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT and n2subscript𝑛2n_{2}italic_n start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT, respectively, and the thickness of the sample is hhitalic_h. Let us consider two different light paths: first one through the middle of the hom*ogeneous microsphere (optical path length OPL1=n1hsubscriptOPL1subscript𝑛1\text{OPL}_{1}=n_{1}\cdot hOPL start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT = italic_n start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT ⋅ italic_h) and the second one through the medium only (OPL2=n2hsubscriptOPL2subscript𝑛2\text{OPL}_{2}=n_{2}\cdot hOPL start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT = italic_n start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT ⋅ italic_h). Given that n1<n2subscript𝑛1subscript𝑛2n_{1}<n_{2}italic_n start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT < italic_n start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT, OPL1<OPL2subscriptOPL1subscriptOPL2\text{OPL}_{1}<\text{OPL}_{2}OPL start_POSTSUBSCRIPT 1 end_POSTSUBSCRIPT < OPL start_POSTSUBSCRIPT 2 end_POSTSUBSCRIPT, and thus the light that travels the first path will reach the detector earlier than light travelling the second path, resulting in temporal delay ΔtΔ𝑡\Delta troman_Δ italic_t. It should be noted that ΔtΔ𝑡\Delta troman_Δ italic_t has to be higher than axial temporal resolution of the qtOCT to receive separation of the two signals in the reconstruction, as presented in Fig.2C (magenta and blue lines in the OCT reconstruction). Meeting this condition allows to retrieve integrated phase image separately for each signal. This opens up possibilities to perform qtOCT measurements in highly scattering conditions and retrieve integrated phase with discarded multiply scattered photons. Technically, during qtOCT processing this is realized by controlling the value of ΔtΔ𝑡\Delta troman_Δ italic_t region that is selected to be inverse Fourier-transformed (Fig. 2A). A specific ΔtΔ𝑡\Delta troman_Δ italic_t region results in reconstructed integrated phase that takes into account information carried by all the photons that reach the detector within ΔtΔ𝑡\Delta troman_Δ italic_t. The smallest ΔtΔ𝑡\Delta troman_Δ italic_t that can be selected is 1 reconstruction plane and due to the fact that 1D FT of 1 sample is the sample itself, in that specific case no IFT is required.

3 Results and discussion

In this section, the results of two experiments emphasizing different properties of qtOCT approach are presented. Experiment 1 (Fig.3) demonstrates the capability of qtOCT to retrieve quantitative integrated phase, while Experiment 2 (Fig.4) presents temporal gating property of the proposed approach. In both cases, the t-FF-SS-OCT data was recorded for 751751751751 wavenumbers using t-FF-SS-OCT system (see Section 2.1). Sample size in OCT reconstructions along t-axis equals theoretical temporal resolution, 31fstimes31femtosecond31\text{\,}\mathrm{fs}start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG.

Starting with the Experiment 1 (Fig.3), the investigated specimen was poly(methyl methacrylate) (PMMA) microsphere (microParticles GmbH, RI839nm=1.4841subscriptRI839nm1.4841\text{RI}_{839\text{nm}}=1.4841RI start_POSTSUBSCRIPT 839 nm end_POSTSUBSCRIPT = 1.4841) with a diameter of 23.5µmtimes23.5micrometer23.5\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 23.5 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG immersed in Zeiss Immersol 518518518518F oil (RI839nm=1.5067subscriptRI839nm1.5067\text{RI}_{839\text{nm}}=1.5067RI start_POSTSUBSCRIPT 839 nm end_POSTSUBSCRIPT = 1.5067). The tOCT signal obtained for this sample was recorded in off-axis configuration to allow direct comparison of phase measurements with DHM. However, to prove that spatial carrier frequency is not required for qtOCT phase retrieval, off-axis holograms were transformed to on-axis ones, according to the procedure shown in Fig. S1 in Supplementary Materials. The obtained spectral fringes were then processed according to the qtOCT procedure (Fig.2A). As expected, reference correction increased the resolution of the OCT reconstruction. The strong signal present in reference-corrected OCT reconstruction corresponds to ballistic and weakly scattered photons that reached the detector first. In order to retrieve integrated phase representing the measured microsphere, the captured OCT signal was temporally gated by the smallest possible window Δt=Δ𝑡absent\Delta t=roman_Δ italic_t = 31fstimes31femtosecond31\text{\,}\mathrm{fs}start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG, which is equivalent to a single sample size along t𝑡titalic_t-axis. To assess the correctness of phase retrieval, the obtained qtOCT phase was compared to the phase calculated through Fourier-transform algorithm used in DHM. The comparison was performed using two measures: universal image quality index (Q) [13] and mean squared error (MSE). The resulting Q value of 0.99500.99500.99500.9950 and MSE equaled to 0.00440.00440.00440.0044 reveal nearly perfect similarity between qtOCT and DHM phase, which is further confirmed by 1D cross-sections.

qtOCT: quantitative transmission optical coherence tomography (3)

In Experiment 2 (Fig.4), the investigated sample was reconfigured. Two PMMA microspheres were located between two glass coverslips, each with a thickness of 170µmtimes170micrometer170\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 170 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG, separated by 120µmtimes120micrometer120\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 120 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG thick spacer. In a plane conjugate to the sample plane, another 170µmtimes170micrometer170\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 170 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG thick coverslip (RI839nm=1.5100subscriptRI839nm1.5100\text{RI}_{839\text{nm}}=1.5100RI start_POSTSUBSCRIPT 839 nm end_POSTSUBSCRIPT = 1.5100) was placed in the field of view in such a way that it covers only one microsphere (for simplicity, the conjugate plane was not depicted in Fig.4). The presence of an additional piece of glass resulted in a temporal delay of 289fstimes289femtosecond289\text{\,}\mathrm{fs}start_ARG 289 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG between light beams that propagated through the 2 microspheres. Given the available axial temporal resolution of the t-FF-SS-OCT system (31fstimes31femtosecond31\text{\,}\mathrm{fs}start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG), it is now possible to temporally separate the 2 signals -- this is confirmed by spatially separated signals in the OCT reconstruction in Fig.4. In order to present temporal gating property of qtOCT, the OCT signal was gated using different time windows ΔtΔ𝑡\Delta troman_Δ italic_t. For each case, the 2D integrated phases were calculated.Phase 1 and Phase 2, corresponding to the parts of the sample with and without additional glass, respectively, were obtained by applying a Δt=Δ𝑡absent\Delta t=roman_Δ italic_t = 31fstimes31femtosecond31\text{\,}\mathrm{fs}start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG window to a specific region of the OCT reconstruction (see Fig.4). The obtained phases confirm the correctness of the separation of both signals. In order to retrieve integrated phase information about both microspheres, a wider temporal window is needed that will cover both signals. For this case, we used a time window of Δt=Δ𝑡absent\Delta t=roman_Δ italic_t = 627fstimes627femtosecond627\text{\,}\mathrm{fs}start_ARG 627 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG. From the integrated complex amplitudes received, the one for k=2π/839nm𝑘2𝜋839nmk={2\pi}/839\text{nm}italic_k = 2 italic_π / 839 nm was selected, and the integrated Phase 3 was calculated, which contains integrated phase values of two separated PMMA microstructures. It should be noted, that due to phase discontinuity at the coverslip edge, the two regions in Phase 3 can be treated independently and their background phase values were unified in order to increase image contrast.

qtOCT: quantitative transmission optical coherence tomography (4)

The results derived from both presented experiments confirm two essential features of qtOCT: its quantitative character and ability to gate time-separated t-FF-SS-OCT signals.Temporal gating property of qtOCT, due to transmission configuration, allows for different applications compared to traditional reflection OCT. It does not allow separating object layers. Instead, it can be used to retrieve integrated quantitative phase of a sample surrounded by a highly-scattering medium by filtering out multiply scattered photons which reach the detector later than weakly scattered ones. The efficiency of this approach depends on the width of the time gate, which in turn depends on the wavelength range of the light source. Here a 75nmtimes75nanometer75\text{\,}\mathrm{nm}start_ARG 75 end_ARG start_ARG times end_ARG start_ARG roman_nm end_ARG-range swept-source was used, which allows for a minimal time gate of 31fstimes31femtosecond31\text{\,}\mathrm{fs}start_ARG 31 end_ARG start_ARG times end_ARG start_ARG roman_fs end_ARG. This time gate is equivalent to approx. 9µmtimes9micrometer9\text{\,}\mathrm{\SIUnitSymbolMicro m}start_ARG 9 end_ARG start_ARG times end_ARG start_ARG roman_µ roman_m end_ARG OPD between 2 signals in air.

4 Conclusions

In this paper a novel qtOCT approach for processing t-FF-SS-OCT data has been presented. The qtOCT is a quantitative method which, as opposed to the already developed tOCT approaches, is capable of recovering 2D integrated phase information in a direct and easy way. We proved high consistency between the phases obtained through our method and DHM. Additionally, we showed that qtOCT can select and analyze time-separated t-FF-SS-OCT signals (in contrast to depth-dependent signals in reflection OCT). This feature makes qtOCT more powerful than other quantitative phase imaging techniques, like DHM and gives the possibility to reduce multiply scattering samples to weakly scattering ones. With ongoing progress of swept-source lasers technology, this approach may allow very precise separation of the incoming photons. Finally, unlike holoscopy or holography-based methods, qtOCT takes advantage of cross-correlation in the signal processing (which enhances the object wave by its interaction with the strong reference wave), which may potentially allow measuring thick or partially absorbing samples that attenuate the object beam.

5 Backmatter

\bmsection

FundingResearch was funded by POB Photonics of Warsaw University of Technology within the Excellence Initiative: Research University (IDUB) programme and REVEAL project (101016726) from European Union’s Horizon 2020 programme.

\bmsection

Author contributionsWK: Conceptualization, Methodology, Formal analysis, Writing - original draft, Supervision; MM: Methodology, Software, Formal analysis, Writing - original draft, Visualization; AK: Investigation - performing the experiments, Data collection, Writing - description and illustration of the optical system;

\bmsection

DisclosuresThe authors declare no conflicts of interest.

\bmsection

Data availability Data underlying the results presented in this paper are available in Zenodo repository (https://doi.org/10.5281/zenodo.11124973).

\bmsection

Supplemental documentSee Supplement 1 for supporting content.

References

  • [1]W.Drexler and J.G. Fujimoto, Optical coherence tomography: technology and applications (Springer Science & Business Media, 2008).
  • [2]A.Trull, J.Van DerHorst, J.Bijster, and J.Kalkman, “Transmission optical coherence tomography based measurement of optical material properties,” \JournalTitleOptics Express 23, 33550–33563 (2015).
  • [3]B.C. Wilson and S.L. Jacques, “Optical reflectance and transmittance of tissues: principles and applications,” \JournalTitleIEEE Journal of Quantum electronics 26, 2186–2199 (1990).
  • [4]K.C. Zhou, R.Qian, A.-H. Dhalla, etal., “Unified k-space theory of optical coherence tomography,” \JournalTitleAdvances in Optics and Photonics 13, 462–514 (2021).
  • [5]J.vander Horst, A.K. Trull, and J.Kalkman, “Deep-tissue label-free quantitative optical tomography,” \JournalTitleOptica 7, 1682–1689 (2020).
  • [6]Y.Wang and R.K. Wang, “High-resolution computed tomography of refractive index distribution by transillumination low-coherence interferometry,” \JournalTitleOptics letters 35, 91–93 (2010).
  • [7]A.Zvyagin, P.Blazkiewicz, and J.Vintrou, “Image reconstruction in full-field fourier-domain optical coherence tomography,” \JournalTitleJournal of Optics A: Pure and Applied Optics 7, 350 (2005).
  • [8]M.Potcoava and M.Kim, “Optical tomography for biomedical applications by digital interference holography,” \JournalTitleMeasurement Science and Technology 19, 074010 (2008).
  • [9]T.Anna, V.Srivastava, C.Shakher, and D.S. Mehta, “Transmission mode full-field swept-source optical coherence tomography for simultaneous amplitude and quantitative phase imaging of transparent objects,” \JournalTitleIEEE Photonics Technology Letters 23, 899–901 (2011).
  • [10]B.Kemper and G.VonBally, “Digital holographic microscopy for live cell applications and technical inspection,” \JournalTitleApplied optics 47, A52–A61 (2008).
  • [11]M.Wojtkowski, Obrazowanie za pomocą tomografii optycznej OCT z detekcją fourierowską (Wydawnictwo Naukowe Uniwersytetu Mikołaja Kopernika, 2009).
  • [12]M.A. Herráez, D.R. Burton, M.J. Lalor, and M.A. Gdeisat, “Fast two-dimensional phase-unwrapping algorithm based on sorting by reliability following a noncontinuous path,” \JournalTitleApplied Optics 41, 7437–7444 (2002). Publisher: Optica Publishing Group.
  • [13]Z.Wang and A.Bovik, “A universal image quality index,” \JournalTitleIEEE Signal Processing Letters 9, 81–84 (2002). Conference Name: IEEE Signal Processing Letters.

SUPPLEMENTAL MATERIAL: Generation of on-axis holograms from off-axis holograms

The t-FF-SS-OCT (full-field swept-source OCT in transmission) system used in our work (see Method Section) operates in off-axis configuration, which allows for simultaneous retrieval of the phase with digital holography microscope (DHM) approach. The qtOCT approach, however, does not require spatial carrier frequency and can operate with on-axis geometry. To prove this, the off-axis t-FF-SS-OCT holograms were numerically transformed to on-axis t-FF-SS-OCT holograms. This procedure is depicted in Fig.5. At first, 2D Fourier transform (FT) was calculated for each XY plane obtained for different wavenumbers k𝑘kitalic_k. In the resulting Fourier spectra 3 separated diffraction orders (+11+1+ 1, 00, and 11-1- 1 order) are visible, which is typical in the off-axis holography. Next, all diffraction orders have been isolated separately for each wavenumber using three main steps: 1) determination of the diffraction order diameter, 2) isolation of the diffraction order with circular Tukey window, 3) centering isolated region. Then, for each wavenumber, 3 isolated diffraction orders were summed, resulting in on-axis Fourier spectra. After that, the 2D inverse Fourier transform was calculated separately for each wavenumber followed by conversion of complex amplitudes to intensity images. Finally, all resulting images were stacked in a 3D matrix.

qtOCT: quantitative transmission optical coherence tomography (5)
qtOCT: quantitative transmission optical coherence tomography (2024)
Top Articles
Latest Posts
Article information

Author: Velia Krajcik

Last Updated:

Views: 5987

Rating: 4.3 / 5 (54 voted)

Reviews: 93% of readers found this page helpful

Author information

Name: Velia Krajcik

Birthday: 1996-07-27

Address: 520 Balistreri Mount, South Armand, OR 60528

Phone: +466880739437

Job: Future Retail Associate

Hobby: Polo, Scouting, Worldbuilding, Cosplaying, Photography, Rowing, Nordic skating

Introduction: My name is Velia Krajcik, I am a handsome, clean, lucky, gleaming, magnificent, proud, glorious person who loves writing and wants to share my knowledge and understanding with you.